In temperature calibration technology, you can find two methods of calibration ? the comparative calibration and the fixed point calibration. The technique of calibration to be used will be defined through the demands on the accuracy and the sort of test instrument. The comparative method represents probably the most widely used approach to calibration.
Just how do the comparative calibration and fixed point calibration methods differ? Using the example of calibrating a resistance thermometer, this website article explains both methods and therefore tries to mention to the reader an improved understanding.
The comparative calibration
Because the name already implies, this kind of calibration is carried out by comparing the resistance thermometer being tested with a higher-quality reference thermometer. The higher-quality reference should thus exhibit an accuracy of three or four times greater than the thermometer being tested. Both thermometers are put into a single homogenous temperature source. When placing ไดอะแฟรม , care should be taken to ensure a little distance between them and that the sensitive points (meaning the measuring points) are in exactly the same height. As a temperature source, liquid baths or dry-well calibrators are usually used.
The fixed point calibration
The fixed point calibration, as opposed to the comparative calibration, is completed contrary to the defined triple, freezing and melting points from the International Temperature Scale of 1990 (ITS-90). The fixed points define temperatures of which the physical state of substances alter. The most important fixed point of the ITS-90 is the triple point of water. At a temperature of 0.010 �C the physical state of high-purity water is continually changing between steam, ice and liquid water. The levels of the physical states always stay in proportion. Generally, only national laboratories or laboratories qualifying for really small measurement uncertainties work with these fixed-point cells, as they are costly and specialised only for this application. Highly stable PRTs (platinum resistance thermometers), which will be used as reference thermometers, are calibrated using the fixed point method. The reference sensor is inserted into a closed, or also open fixed-point cell, which holds the temperature constant over an extended period.
Example: fixed points from -189 to 660 �C in accordance with ITS-90:
Fixed point
Temperature
Triple point of argon
-189,3442 �C
Triple point of mercury
-38,8344 �C
Triple point of water
0,01 �C
Melting point of gallium
29,7646 �C
Freezing point of indium
156,5985 �C
Freezing point of tin
231,928 �C
Freezing point of zinc
419,527 �C
Freezing point of aluminium
660,323 �C
Note
More info on our Services ?Temperature calibration? are available on the WIKA website.